Beckman-Coulter Moflo-XDPtm Item #4104 |
|
Beckman-Coulter MoFlo-XDP: 3-laser, 12-color sorter with Summit ver. V5.3, Intelli-Sort 1, Smart Sampler, CyClone deposition. Custom configurations available.
|
|
|
Lasers: 200 mw 488nm, 100mw 405nm and 150mw 637nm. Custom configurations available |
Workstation: PC with 2 color monitors and printer. |
Software: Summit Software Ver. 5.3. Microsoft Windows XP Professional. More than 1 billion event listmode files. Powerful data handling. |
Compensation: 18 x 18 intra-laser compensation. |
Processing Speed: More than 100,000 analysis events per second. |
Electronics: Digital triggering on any parameter, digital pulse processing, true 32-bit-high resolution, 5-decade multi-channel digital system. Log, height, area, width, log area for each parameter. |
Optics: Open architecture for configuration modularity. Optical bench can accommodate any laser used in flow cytometry. Fiber-coupled lasers with Beam Shaping Optics for stability and ease of alignment. 2 scatter and up to 18 fluorescence capability. Custom configurations available |
Purity: > 99% purity at all data rates. High recovery, yield and viability. |
Fluorescence Sensitivity:< 150 MESF FITC, <100 MESF PE |
Particle Resolution: < 0.2 μm to 25 μm. |
Sorting: More than 70,000 sort events per second. Independent Sort Mode capability in each tube. IntelliSort I: Beadless drop delay determination and monitoring. Unique nozzle design gives high viability and recovery even at the highest speeds. Upgrade to Intelli-Sort II available. |
Cell Deposition: High precision CyClone can sort into 96, 284, multiwall trays or microscope slides. |
Fluidics: Easily replaceable probe and sample line (for biosafety). Maintains the critical stable droplet formation required in sorting |
Nozzles: 2X 70u, 1 X 100u and 1 X 150u nozzle tips. |
System Pressure: 4-100 PSI. |
Manufacturing Date: 2011 |
|
|
Custom laser, detector and workstation configurations can be quoted separately.
|
|
|
Determine condition of cytometer hardware, computer, peripherals, cables and tubing. Replace boards, cables or computer as needed to get unit powered up for testing. Run bleach solution through fluidics to decontaminate. Run pre-final test software to determine what electronic circuit boards may be malfunctioning and replace. Clean all electronic board connections and inspect all wiring for wear. Determine the condition of all lasers. If lasers read below specified power and stability, would replace with new or remanufactured laser of equal or better specifications. Change all worn fluidics components. Change tubing as needed and replace all filters. Adjust power supplies to specifications. Replace fluidics containers if cracked or discolored. Perform sample and sheath sensor pneumatics calibration.
Perform beamshaper alignment.
Clean all optical dichroic and band-pass filters. Align PMTs for highest gain. Replace any PMT that requires excessive voltage or high HPCVs. Perform optical alignment for optimum HPCV's using Flow Check beads. Confirm CyClone operation on microplates. Confirm drop delay and other sorting parameters are functioning propertly. Provide any required restoration of exterior and interior components. Confirm that instrument meets all original factory specifications. |
|
|
Installation by a qualified service representative quoted separately. The price will include all travel expenses and labor required to confirm proper operation at the customer's facility utilizing standard calibration procedures. Any additional time required to correct malfunctions shall be included in the quoted installation fee. |
|
|
Instrument and/or applications training by a qualified representative are available for an additional fee. The quoted price includes travel and perdiem expenses when training is performed during an installation visit. Otherwise additional reimbursement for travel and perdiem expenses may be required. |
|
|
Custom warranty or service contracts can be quoted separately. All components are covered unless failure is due to operator negligence. In addition, we offer unlimited telephone support for all of the products sold through Spectron Corporation regardless of warranty. |
|
|
Shipping costs will be quoted separately. If the purchaser becomes the shipping agent for one of the products located in the State of Washington they could be subject to paying Washington State sales tax unless exempt. No such obligation exists if Spectron Corporation is the designated shipper to destinations outside the State of Washington. This equipment shall be packed in a sturdy double-walled box surrounded inside by 2" foam sheets on all sides. All components shall be individually protected within the same container. The box will be taped, strapped and attached to a pallet with additional strapping. Shipping will be arranged through a qualified airfreight carrier with full insurance. |
|
|
Most established institutions will be approved for net-30 terms or other payment plans. Leasing or rental programs may also be available for qualified buyers with various options for an eventual purchase. Washington State customers or others arranging shipment or picking up products themselves may be subject to Washington State sales tax unless exempt. Spectron Corporation will inform prospective buyers of any such circumstance that could result in a tax liability. Otherwise, buyers would only be subject to applicable use taxes that they might be required to pay directly to their respective state. |
|
|
Spectron Corporation would appreciate the opportunity to bid on a service contract covering our products as well as other flow cytometers located at your facility. Please inquire into any modifications to the listed products that would better suit your requirements. We strive for complete satisfaction. |
|